ASTM E387 - 04(2022)

Standard Test Method for Estimating Stray Radiant Power Ratio of Dispersive Spectrophotometers by the Opaque Filter Method

Warning: ASTM standards for reading are viewed through the client area. Thefilescannot be downloaded.
Edition date: 2022-11-01
In Force
Available languages: English
Summary:

Keywords: molecular spectroscopy; spectrophotometry; SRP; SRPR; stray light; stray radiant power; stray radiant power ratio;
Scope:

1.1 Stray radiant power (SRP) can be a significant source of error in spectrophotometric measurements, and the danger that such error exists is enhanced because its presence often is not suspected (1-4).2 This test method affords an estimate of the relative radiant power, that is, the Stray Radiant Power Ratio (SRPR), at wavelengths remote from those of the nominal bandpass transmitted through the monochromator of an absorption spectrophotometer. Test-filter materials are described that discriminate between the desired wavelengths and those that contribute most to SRP for conventional commercial spectrophotometers used in the ultraviolet, the visible, the near infrared, and the mid-infrared ranges. These procedures apply to instruments of conventional design, with usual sources, detectors, including array detectors, and optical arrangements. The vacuum ultraviolet and the far infrared present special problems that are not discussed herein.

Note 1: Research (3) has shown that particular care must be exercised in testing grating spectrophotometers that use moderately narrow bandpass SRP-blocking filters. Accurate calibration of the wavelength scale is critical when testing such instruments. Refer to Practice E275.

1.2 These procedures are neither all-inclusive nor infallible. Because of the nature of readily available filter materials, with a few exceptions, the procedures are insensitive to SRP of very short wavelengths in the ultraviolet, or of lower frequencies in the infrared. Sharp cutoff longpass filters are available for testing for shorter wavelength SRP in the visible and the near infrared, and sharp cutoff shortpass filters are available for testing at longer visible wavelengths. The procedures are not necessarily valid for “spike” SRP nor for “nearby SRP.” (See Annexes for general discussion and definitions of these terms.) However, they are adequate in most cases and for typical applications. They do cover instruments using prisms or gratings in either single or double monochromators, and with single and double beam instruments.

Note 2: Instruments with array detectors are inherently prone to having higher levels of SRP. See Annexes for the use of filters to reduce SRP.

1.3 The proportion of SRP (that is, SRPR) encountered with a well-designed monochromator, used in a favorable spectral region, typically is 0.1 % transmittance or better, and with a double monochromator it can be less than 1×10-6, even with a broadband continuum source. Under these conditions, it may be difficult to do more than determine that it falls below a certain level. Because SRP test filters always absorb some of the SRP, and may absorb an appreciable amount if the specified measurement wavelength is not very close to the cutoff wavelength of the SRP filter, this test method underestimates the true SRPR. However, actual measurement sometimes requires special techniques and instrument operating conditions that are not typical of those occurring during use. When absorption measurements with continuum sources are being made, it can be that, owing to the effect of slit width on SRP in a double monochromator, these test procedures may offset in some degree the effect of absorption by the SRP filter; that is, because larger slit widths than normal might be used to admit enough energy to the monochromator to permit evaluation of the SRP, the stray proportion indicated could be greater than would normally be encountered in use (but the net effect is still more likely to be an underestimation of the true SRPR). Whether the indicated SRPR equals or differs from the normal-use value depends on how much the SRP is increased with the wider slits and on how much of the SRP is absorbed by the SRP filter. What must be accepted is that the numerical value obtained for the SRPR is a characteristic of the particular test conditions as well as of the performance of the instrument in normal use. It is an indication of whether high absorbance measurements of a sample are more or less likely to be biased by SRP in the neighborhood of the analytical wavelength where the sample test determination is made.

1.4 The principal reason for a test procedure that is not exactly representative of normal operation is that the effects of SRP are “magnified” in sample measurements at high absorbance. It might be necessary to increase sensitivity in some way during the test in order to evaluate the SRP adequately. This can be accomplished by increasing slit width and so obtaining sufficient energy to allow meaningful measurement of the SRP after the monochromatic energy has been removed by the SRP filter. However, some instruments automatically increase sensitivity by increasing dynode voltages of the photomultiplier detector. This is particularly true of high-end double monochromator instruments in their ultraviolet and visible ranges. A further reason for increasing energy or sensitivity can be that many instruments have only absorbance scales, which obviously do not extend to zero transmittance. Even a SRP-proportion as large as 1 % may fall outside the measurement range.

Note 3: Instruments that have built-in optical attenuators to balance sample absorption may make relatively inaccurate measurements below 10 % transmittance, because of poor attenuator linearity. The spectrophotometer manufacturer should be consulted on how to calibrate transmittance of the attenuator at such lower level of transmittance.

1.5 High accuracy in SRP measurement is not always required; a measurement reliable within 10 or 20 % may be sufficient. However, regulatory requirements, or the needs of a particular analysis, may require much higher accuracy. Painstaking measurements are always desirable.

1.6 The values stated in SI units are to be regarded as standard. No other units of measurement are included in this standard.

1.7 This standard does not purport to address all of the safety concerns, if any, associated with its use. It is the responsibility of the user of this standard to establish appropriate safety, health, and environmental practices and determine the applicability of regulatory limitations prior to use.

1.8 This international standard was developed in accordance with internationally recognized principles on standardization established in the Decision on Principles for the Development of International Standards, Guides and Recommendations issued by the World Trade Organization Technical Barriers to Trade (TBT) Committee.

ICS: 17.180.30-Optical measuring instruments
CTN: E13.01 - E13.01

Standards Cancellations

Anula a E0387-04R14

Otras Relaciones

Es referenciada por E2056 - 04(2024)

Es referenciada por E0932-89R21

Es referenciada por E0275-08R22

Es referenciada por E0925-09R22

Es referenciada por E0169-16R22

Es referenciada por E1341-16R20

Es referenciada por E1164 - 23

The book in the author's words

Ultricies magna feugiat malesuada sociosqu varius vivamus cubilia parturient, himenaeos vitae vehicula nam placerat netus urna platea, nostra rutrum felis mattis penatibus velit quisque.

Button
Frequently Asked Questions Do you have any questions about our products?
  • Standards UNE, EN, ISO, IEC, BSI, DIN, ASTM, AFNOR, IEEE, SAE
  • In addition, you can request the rules of the rest of the organizations through the e-mail normas@aenor.com
  • Technical books on paper and in electronic format (PDF, epub).

The standards can be purchased in PDF, reading or paper. The reading standards are not download files, they can only be viewed in the client area. The standards ordered on paper and some of the books in the catalogue are printed on demand. 

Check deadlines in normas@aenor.com.

The license of use is for one user and one device, if you want to reproduce the content of the standard, you must request a license that will have an additional cost. Send us your inquiry here 

The AENOR standards and books that appear in the online store can only be purchased exclusively through the website. AENOR does not have a physical store.

Purchase procedure: by clicking on "Buy" the desired products will go to the shopping cart. If there are display problems, the recommended browser is Chrome.

To formalize the purchase you must access the customer area. If you are not registered as a customer, you must fill in a form with the data along with a password and username. This will create the account.

Once the "Customer data" form has been completed, "Order in progress" will be displayed with all the items loaded in the shopping cart, their prices, taxes established in current legislation and shipping costs if applicable.

The prices of the standards and books that appear in the various sections do not include taxes or shipping costs.

AENOR promotional codes consist of alphanumeric characters and can only be applied to online purchases, received through a specific offer and for a limited time. To apply your promotional code, you just have to enter it in step 2 of 4 of the purchase process on the website and click on "apply", after you have identified yourself and chosen the payment methods. Promo codes are not cumulative.

 

  • Credit or debit card (Visa, Mastercard) and PayPal.
  • Bank transfer. If you opt for this form of payment, you must first send AENOR a copy of the transfer by email to normas@aenor.com
  • The purchase invoice can be downloaded from the customer area, in my previous orders

In the case of clients of companies based abroad, the taxpayer identification number of the corresponding country (for example, in Argentina the CUIT), must be filled in the CIF/NIF - VAT field .

  • Direct download via the website in the Customer Area. In the customer area, which can only be accessed with a password and username, the products purchased will be available for a period of fifteen days from the date of purchase, as long as the payment has been accepted. Files in digital format are protected and in no case editable. Before purchasing them, it is important that the license of use is read and accepted as a prior step to purchase.
  • Shipping by courier. Products purchased on physical media are shipped by courier. The maximum delivery time in Spanish territory, from the acceptance of the order by AENOR, is:
  •  Approximately seven working days for all standards purchased through the store in paper format.
  • Approximately three days for books purchased through the store. Stocks of paper books are limited and their offer on the website does not imply availability within the indicated period. In the event that the requested book is not available, the customer is notified of the delay in receiving the order, which will be approximately seven working days. 

For the rest of the products that are not on the website, check availability and delivery time at normas@aenor.com.

1. For digital products (PDF, Epub), once delivery has been made by direct download via the website in the Customer Area, you will not have the right to exercise your right of withdrawal.

2. For personalised products on paper, once the purchase has been made, you will not have the right to exercise your right of withdrawal.

3.  For all other paper products, you have the right to withdraw from the sale within 14 calendar days from the date of purchase. Remember that for the return it is essential that the product is in perfect condition, sealed by the packaging and preserving its original packaging. The customer will be responsible for pickup and shipping costs.

The order invoice includes shipping costs, so there is no amount to pay to the courier. Shipping costs are calculated based on both the final destination of the order and the number of products ordered. They include transport and packaging costs. Shipping costs are subject to periodic revisions. Outlet books will have free shipping costs only if the shipment is made in the Peninsula.

Destination Up to three standards and/or publications From three standards and/or publications
Peninsula 7,31€ 8,60€
Balearic Islands 18,04€ 23,34€
Canary Islands, Ceuta and Melilla  18,04€ 23,34€
Europe 59,17€ 80,07€
United States and Canada 70,07€ 96,94€
Rest of the world 91,94€ 115,91€
  • Purchases made by residents of the Member States of the European Union will be subject to the payment of VAT (value added tax).
  • ​​
  • In the case of legal persons and natural persons who, acting as entrepreneurs, are domiciled in a Member State of the European Union (except residents in Spain) and have an intra-community NIF/VAT registered in the VIES census, they will be exempt from paying VAT, being an essential condition the sending of this document by email to normas@aenor.com.
  • Purchases made in a private capacity (natural person), regardless of where they have their residence, will be subject to the payment of VAT.
  • Purchases made by entities in non-EU countries will be exempt from paying VAT, as long as they send the corresponding tax residence document by email to normas@aenor.com.
  • The sale operations will be understood to have been carried out at AENOR's registered office: Génova 6, 28004, Madrid – Spain. 

The contract for the purchase of products through this Website shall be governed by Spanish law. Any dispute arising out of or in connection with the use of the Website or such contract shall be subject to the exclusive jurisdiction of the Courts and Tribunals of Madrid.

Notwithstanding the foregoing, if you are entering into this contract as a consumer under the terms of Royal Decree 1/2007, nothing in this clause shall affect the rights that may be granted to you as such under applicable law.